The first step in performing quality assurance (QA) on the output of the Data Ingest and GCP Collection module is to scan the GCP Residuals Summary report for anomalies: images that contain particularly high residuals or a small number of collected GCPs.
To review the collected GCPs
In Figure 1, the report shows several ingested images that have a low number of collected GCPs, and a very high check-point residual. Although there may be several reasons for this, such as a heavy presence of clouds in the imagery combined with insufficient chip coverage for the region, the affected images merit closer inspection.

For example, for image 522247-S4H1070709191133M_RAW_PAN.pix shown in Figure 1, open the CATALYST Professional OrthoEngine project file:
522247-S4H1070709191133M_RAW_PAN_OE.prj
Compare each GCP with its corresponding point (either in a chip database, reference image, or other). The GCPs should align well with the reference source.
For example, Figure 2 shows the GCP Collection window for the sample image 52693431001240357022J_RAW_MS.pix, indicating that the image contains two collected GCPs.

For example, point E209_9659_005 is shown in Figure 3.

© PCI Geomatics Enterprises, Inc.®, 2026. All rights reserved.